Simulated Field Evaluation of Six Techniques for Controlling the

Drywood Termite Incisitermes minor (Isoptera: Kalotermitidae)

in Residences

 
 

VERNARD R. LEWIS and MICHAEL I. HAVERTY

 
 
1 Department of Environmental Sciences, Policy & Management, Division of
Insect Biology, 201 Wellman Hall, University of California,
Berkeley, CA 94720
 
2 Pacific Southwest Research Station, Forest Service, U.S. Department of
Agriculture, P.O. Box 245, Berkeley, CA 94701
 
3 This document reports the results of reseach only. Mention of
proprietary products or techniques does not constitue an
endorsement or a recommendation for its use by the University of
California or USDA.

 
ABSTRACT Nonchemical and chemical methods for control of drywood
termites were evaluated under simulated field conditions. Specifically, we
assessed the efficacy of four methods currently marketed as alternatives to
whole-structure fumigation for control of drywood termites: excessive heat,
excessive cold, electrocution, and microwaves. In addition, we evaluated a
reduced dosage of methyl bromide synergized with carbon dioxide, as well as
a standard fumigation treatment with sulfuryl fluoride.
Tests were conducted using Incisitermes minor (Hagen) in
artificially infested or naturally infested boards of various dimensions
used in construction. Infested boards were placed into the attic,
drywalls, or subarea of the Villa Termiti, a symmetrical building
constructed specifically for these tests. Commercial pest control
operators performed 5 of the 6 control methods; liquid nitrogen was applied
by University of California personnel. For artificially infested boards,
mortality was measured 3-d and 4-wk post-treatment. For naturally infested
boards, mortality was evaluated only 4-wk post-treatment. Efficacy
performance of all treatments was compared to 90, 95, and 99 percent levels
of mortality.
Termite mortality in artificial boards was 100 percent at 3-d and
4-wk post-treatment for both fumigant gases. Heating the whole-structure
or spot-applications using microwaves resulted in 96 and 90 percent
mortality, respectively, 3-d post-treatment. Mortality levels 4-wk
post-treatment increased to 98 percent for heating and 92 percent for
microwaves. Spot-applications of liquid nitrogen at the 30-min@1.4 kg/min
dose (highest dose tested) achieved 100 percent mortality 3-d
post-treatment. However, for the 15-min@ 0.9 kg/min and 7-min@ 0.9 kg/min
dosages, 4-wk post-treatment mortality levels were 99 percent and 87
percent, respectively. Mortality by electrocution of termites in
artificially infested boards was 44 percent 3-d post-treatment in the first
test. Four-weeks post-treatment drywood termite mortality increased to 82
percent. In a second electrocution test, using spot-application techniques
infrequently used in structures, mortality levels increased to 93 percent
3-d and 98 percent 4-wk post-treatment.
For naturally infested boards, both fumigants exceeded the 99
percent level of mortality. Nonchemical applications of heat for
whole-structure and spot-applications with microwaves resulted in 100
percent and 99 percent mortality levels for naturally infested boards.
Chemical applications of liquid nitrogen were at or near 100 percent for
naturally infested boards tested at the 30-min@1.4 kg/min and 15-min@0.9
kg/min dosages. However, mortality was significantly lower (74 percent)
for the 7-min@0.9 kg/min dose. Mortality levels from electrocution were 89
percent and 95 percent 4-wk post-treatment, respectively, in the two tests.
The distribution of termite survivors varied for some techniques by
: 1) location within the test structure and 2) galleries within test
boards. Visual signs of damage to test boards, drywall, and the Villa
Termiti were noted for some treatment techniques. This study provides
information for evaluation of the relative efficacy of nonchemical
alternatives and fumigation technology for the eradication/elimination of
drywood termite infestations in structures.
 
THE DAMAGE CAUSED by wood-destroying insects results in a significant
economic impact on many structures throughout the United States.
Nationwide, the cost for wood-destroying insect control and repairs of
damage approaches $5 billion per year; the outlay in California and Hawaii
alone probably exceeds $1 billion per year (Su & Scheffrahn 1990, Brier et
al. 1988). In California, a breakdown of these expenditures by insect
species reveals that subterranean termites, primarily Reticulitermes
species, and drywood termites, specifically the western drywood termite,
Incisitermes minor (Hagen), are responsible for over 95 percent of all
costs due to wood-destroying insects (Rust et al. 1988, Brier 1987).
Damage attributed to wood-boring beetles and carpenter ants amounts to
about 3 percent of the total cost (Rust et al. 1988, Brier 1987).
Subterranean termites cause problems throughout California; however, damage
by drywood termites is more common in the southern portion of the state
(Wilcox 1979). According to Wilcox (1979), over 70 percent of all
inspection reports from Los Angeles and San Diego Counties submitted to the
California Structural Pest Control Board from 1976 to 1977 indicate the
presence of damage by drywood termites. Infestations in northern
California, including the San Francisco Bay Area, the Sacramento and San
Joaquin Valleys, appear to be increasing (Lewis & Haverty, unpublished
observations). Kofoid (1934) stated that I. minor exhibited a preference
to infest rafters, roof sheathing and southern exposures of dwellings in
the northern part of its range. In southern or desert areas, infestations
are more likely to be found lower in the structure.
For many years, the standard treatment for elimination of drywood
termite infestations was fumigation with either methyl bromide or sulfuryl
fluoride. The use of fumigants is considered a "whole-structure treatment"
(for treating simultaneously all wooden members and extensive or difficult
to access infestations in structures) (Scheffrahn & Su 1994). When
properly applied, these toxic gases are effective in eliminating
infestations of drywood termites throughout the treated structure. Both
gases are highly toxic biocides that kill termites and other organisms by
disruption of biochemical pathways. Specifically, these fumigants cause
cessation of lipid catabolism and glycolysis (Meikle et al. 1963, Su &
Scheffrahn 1986). For methyl bromide, symptomology includes the darkening
of unpigmented appendages (Scheffrahn & Su 1992).
Several studies have demonstrated the effectiveness of chemical
fumigants against a variety of termite species (Bess & Ota 1960, Osbrink et
al. 1987); however, relatively little information has been presented on the
effectiveness of fumigation on an operational basis. Ebeling and Wagner
(1964) found that 26 to 37 percent of structures in Los Angeles that had
been fumigated with methyl bromide showed evidence of active drywood
termite infestations within 3 to 5 years. In the same study, they found
comparable rates of "re-infestation" for drill-and-pin applications,
ranging from 63 to 79 percent. We suggest, however, that the
"re-infestation" rate of this insecticidal spot-treatment includes a
significant proportion of termite infestations that were never eliminated.
The public is showing increased interest in nonchemical or "least
toxic" approaches to insect control. In a survey in Indiana, 87 percent of
the respondents claimed to have attempted a nonchemical method for control
of household insect pests (Bennett et al. 1983). In a similar survey, 72
percent of the respondents in Berkeley, California, said that they had
personally used nonchemical control techniques in their homes (Levenson &
Frankie 1983). A more telling statistic from this study is that 67 percent
of respondents, representing three different geographic locations of the
United States (Berkeley, California; Dallas, Texas; and New Brunswick, New
Jersey), said they were increasingly cautious about the use of pesticides.
Closely paralleling the public's interest in "Urban Environmentalism" in
California is the development and commercialization of nonchemical
alternatives directed against wood-destroying insects. The list of these
control techniques presently marketed in California for control of drywood
termites is growing and currently includes excessive heat, excessive cold,
electrocution, and microwaves (also based on temperature elevation).
Excessive cold, electrocution, and microwaves are "spot or localized"
treatment methods (treatment often restricted to a single spot within a
board or small group of boards). Whole-structure heating of homes comes
closest to conventional fumigation.
A reason for interest in the effectiveness of alternatives to
fumigation is that ownership of homes change, on average, every three years
in California (Ebeling & Forbes 1988). A usual requirement of home sales
is a guarantee that the home is free of infestations and infections of
wood-destroying organisms. Assurance of pest-free homes, without the use
or overuse of chemical pesticides, is becoming more important in closing
real estate transactions.
There has been limited published research, either in the laboratory
or in the field, on any of the alternative control methods examined in our
study. Forbes and Ebeling (1987) found that nymphs of I. minor died if
exposed to 51o C for more than 10 min. Those results form the basis for
recommendations for heat fumigation of structures. Death from exposure to
excessive heat no doubt has a complex mechanism. Hyperthermia affects
insects at the cellular level, disrupting the function of cell membranes
and stability of enzymes (Bowler 1981, Ebeling 1994).
The effects of low temperatures on termites have scarcely been
investigated. Lund (1962) determined that workers of the eastern
subterranean termite, Reticulitermes flavipes (Kollar), succumbed after
less than 5 min exposure at -9.5o C (14.9o F) to -13.0o C (8.6o F).
Temperate species such as the drywood termite Kalotermes flavicollis (F.),
the Pacific dampwood termite, Zootermopsis angusticollis (Hagen), and a
European subterranean termite, Reticulitermes lucifugus (Rossi) were able
to survive for long periods of time when held below 18o C (64o F) (Becker
1967). Feeding was minimal when the Formosan subterranean, Coptotermes
formosanus Shiraki, and R. flavipes were maintained at 5o C (41o F) and 10o
C (50o F) (Smythe & Williams 1972). At 5o C all termites of both species
died within 8 wk. All C. formosanus died when maintained at 10o C, whereas
R. flavipes survived (Smythe & Williams 1972).
In initial experiments with I. minor, Forbes and Ebeling (1986)
reported that individuals died within 5 min at temperatures between
-18.5 to -19.4o C (-1.3 to 2.9o F). Rust et al. (1995) corroborated the
experiments of Forbes and Ebeling (1986) and found that exposure of workers
(sic) and alates in wooden blocks to temperatures below -21.4o C resulted
in 100 percent mortality. This temperature is apparently below that which
causes the formation of ice crystals in the hemolymph resulting in
disruption of cell membranes and eventual death of the insect (Heinrich
1981). Both studies surmised that chilling wood below that minimum lethal
temperature will result in the elimination of all I. minor present in
timbers.
To date, there has been only one published study evaluating
efficacy of electrocution for control of drywood termites in wood. Ebeling
(1983) used the Electrogun® to treat blocks of wood artificially infested
with nymphs of I. minor. When the probe of this device was placed into a
hole near the gallery in wood containing nymphs and short bursts of
electricity were applied, sparks were seen jumping from termite to termite.
The mode of action for mortality in termites is not known. Even after
exposure to the electric shock, not all termites were killed immediately.
However, within 5 days of treatment, all termites in each "test" were dead.
Ebeling (1983) attributed delayed mortality to the destruction of
intestinal protozoans.
Similarly, when nymphs within galleries were treated by passing the
Electrogun® over the surface of the wood for one minute many termites
survived initially, although efficacy in these tests was equivocal (Ebeling
1983). As with the technique of drilling and inserting the Electrogun®
probe into the gallery, there was delayed mortality as a result of passing
the probe over the surface of the wood. After 5 days, all termites in
treated blocks were dead.
Direct treatment of termites in natural gallery systems for one
minute caused 10 percent mortality (Ebeling 1983). By placing the probe of
the Electrogun® in a "kick-out" hole, direct observation of sparking and
termite mortality demonstrated the fact that galleries in that piece of
wood were interconnected, and the device was effective in killing I. minor
nymphs.
Microwaves have been investigated as a means of destroying insects
in nuts and stored grain (Locatelli & Traversa 1989, D'Ambrosio et al.
1982, Tilton & Vardell 1982a&b, Nelson & Payne 1982, Nelson 1977, Watters
1976, Rosenberg & Bögl 1987) and for preserving textiles and museum
specimens (Hall 1981, 1988, and Philbrick 1984, Regan et al. 1980). Direct
application of microwaves to insects does not affect various life stages of
insects equally (Del Estal et al. 1986).
Microwaves can be used to heat the substrate and then subsequently
kill the infesting insects by extreme temperature (Locatelli & Traversa
1989). Microwaves can also act directly on insects within relatively dry
substrates by agitating water and/or fat molecules. Friction caused by
this agitation creates heat which likely causes death by protein
denaturation and membrane disruption (Hall 1981). Most investigators
measure the effect of the time of exposure on insects and/or substrate,
keeping the power and wavelength constant (Crocker et al. 1987). Thus far,
there have been no published reports on effects of microwaves on termites
either in the laboratory or under field conditions.
Field tests of these alternative control methods are scant. Forbes
& Ebeling (1987) reported on a demonstration of heating a mock-up house
above a critical temperature. Air from an electronically-driven blower was
passed through a gas-fired heater and delivered into the interior of this
"house." This treatment eventually raised the internal temperature of wood
in the crawl space, attic, and wall voids. Their objective was to
determine a relationship between a given room temperature and the time
required to reach lethal temperatures within structural timbers with
various cross-sectional dimensions (3.8 X 8.6, 8.9 X 28.6, or 13.9 X 28.6
cm (2 X 4, 4 X 12, or 6 X 12 inches, respectively)). For nymphs of I.
minor, 100 percent mortality was achieved when the temperature within wood
was maintained at >48o C for at least 30 min.
Forbes and Ebeling (1986) documented a method for chilling infested
structural members below the survival threshold temperature for drywood
termites. They reported that wall voids do not have to be chilled below
-80o C (-112o F) in order to reach temperatures within wooden structural
members that are lethal to drywood termites. To speed the process,
however, liquid nitrogen was used to produce temperatures as low as -180o C
(-292o F) in wall voids. These spaces remained at temperatures lethal to
termites for more than 2 hours. These authors suggested that the use of
strategically placed insulated mats decreased the amount of liquid nitrogen
required to chill the area and prevented frost formation on the walls.
There have been no published reports on the efficacy of the
Electrogun® under actual or simulated field conditions. Ebeling (1983)
reported empirical observations on the efficacy of this device after
routine commercial treatments. Softwood boards in a pile in a shed were
treated by a pest control operator who spent about 5 min treating a 3.8 X
13.9 cm (2 X 6 inches) by approximately 1 m long timber. Mortality from
this treatment was 74 percent immediately after treatment, 81.3 percent
after 26 days, and 96.3 percent after 57 days. It is important to
emphasize that this particular piece of wood was not treated in situ,
rather it "...was placed on a concrete slab and was treated with
Electrogun®, paying special attention to 'thin areas' and 'kick-out' holes."
Ebeling (1983) also examined 35 termite colonies 1-4 mo after
treatment with the Electrogun® by pest control operators. His measure of
efficacy was the appearance of new fecal pellets. Three of the 35 colonies
needed retreatment. In a survey of pest control operators who used the
Electrogun®, nearly all reported fewer call-backs than with previously
employed drill-and-treat localized chemical treatment methods (Ebeling
1983). However, Mampe (1990), citing Ebeling (1983), considered the use of
the Electrogun® to be similar to localized chemical treatments. He
described the use of the Electrogun® as follows: "the operator moves the
gun along wood members, creating an arc of high-voltage which penetrates
the wood." He does not mention penetrating the wood or galleries with the
probe of the Electrogun®. Mampe (1990) claimed that this technique "...has
met with only limited success, but may be useful for isolated infestations."
Nonchemical methods have been proposed as replacements for
structural fumigation for drywood termite control. Several of these
methods are now being applied operationally by pest control companies in
California, Florida, and Hawaii.
The economic impact of wood-destroying insects in our structures
will likely continue to increase significantly into the next century, as
will the demand for protection of wood-in-service. The reasons for this
increase are many including: 1) increased urbanization and population
growth, 2) additional environmental constraints on timber cutting and an
increase in the value of wood in North America, 3) the apparent reduced
efficacy of currently registered soil termiticides, and 4) a growing
concern of the public over the use of toxic chemicals in and around
households. Justification for development of nonchemical control
technology for drywood termites is obvious. However, the public must be
assured that alternatives to fumigation offered to them are efficacious
when properly applied.
Here we report the efficacy test results of two types of fumigation
and four methods currently marketed as alternatives to whole-structure
fumigation. We tested each method against three levels of efficacy: 90,
95, and 99 percent, after 3-d and 4-wk post-treatment. Our purpose in this
research was solely to evaluate the efficacy of each treatment, not to make
direct comparisons among treatments.
Materials and Methods
Insects. In this study, two types of infested material were used:
artificially infested and naturally infested boards. Termites placed in
artificially infested boards were extracted from naturally infested wood
(lumber, firewood, and grape prunings) containing I. minor. Infested wood
was collected from 12 cities in California: Concord, Fremont, Fresno, Los
Angeles, Novato, Oakland, Riverside, Sacramento, San Jose, San Luis Obispo,
San Rafael, and Ventura. Termites were removed from wood using Berlese
funnels or direct extraction. For the Berlese technique, cut sections of
wood were placed into the 0.5-m diameter anterior ends of funnels and
covered with friction-fitting lids containing a 100-watt light bulb. Wood
sections were left in the Berlese funnels overnight; termites fleeing from
the heat were collected at the bottom of funnels in 0.9 l glass Mason® jars
containing a damp paper towel. The method of direct extraction consisted
of using hammers and wood chisels to tap termites out of naturally infested
wood.
Termites removed by either collection technique were put into
hollowed-out birch tongue depressor rearing chambers 1.9 X 15.9 cm
long.(Bess & Ota 1960); the hollowed-out space was 1.2 X 5.0 cm long (Fig.
1). Seven of these tongue depressors were stacked onto each other and held
together with masking tape. The top and bottom of the chamber was sealed
with intact depressors; the entire unit was held together at both ends with
2-cm wide masking tape and a rubber band.
Each tongue depressor's chamber contained approximately 200
termites. Termite groups were from mixed colonies. It has been shown that
drywood termites can be mixed together with minimal mortality (Atkinson,
1994). All rearing chambers were stored in 14.3 X 10.4 X 3.4 cm (length,
width, height; lwh) clear plastic boxes with friction-fitting lids.
Plastic boxes containing the rearing chambers were held in an incubator in
a glass greenhouse for several weeks before use. Environmental conditions,
which ranged from 18 to 37 oC with relative humidities of 40 to 80 percent,
were monitored with a hygrothermograph maintained inside the incubator.
For all treatments, only healthy termites were used. Primarily, we
selected pseudergates of at least the fourth instar; however, some younger
nymphs were occasionally used. Alates and soldiers were not used in the
study.
Villa Termiti. To simulate field conditions, a mock-structure we
call the Villa Termiti, was built specifically for these tests of drywood
termite control methods. The Villa Termiti is a 6.1 X 6.1 m (37.2 m2; 20
by 20 ft or 400 ft2) building constructed of construction-grade Pacific
Douglas fir (Pseudotsuga menziesii (Mirb.) Franco var. menziesii). Walls
were built with 3.6 X 14.5 cm (2 X 6 in) studs on 30.5 cm (12 in) centers.
These studs allowed for easy installation of test boards of varying sizes.
Other dimensional sizes of wood used in construction were 2.0 X 11.6, 3.7 X
8.6, 3.8 X 23.5, 9 X 9, and 8.7 X 23.5 cm (approximately 1 X 5, 2 X 4, 2 X
10, 4 X 4, and 4 X 10 in). No pressure-treated or otherwise
chemically-treated wood was used in this building. Foundation-grade
redwood (Sequoia sempervirens (D. Don) Endl.) mudsill plates were used in
lieu of pressure-treated wood.
The Villa Termiti was designed to be symmetrical with doors and
windows on all four sides (Fig. 2). This symmetry of design allowed for
testing unbiased by construction or aspect, and it enabled internal
replication using an entire wall of wall voids. To better reflect the
different building styles used in northern versus southern California
(wooden exteriors on raised foundations versus stucco exteriors on slabs),
combination exterior walls and foundation were included in the Villa
Termiti design. The Villa Termiti contains an attic, "living space," and a
subarea (Figs. 3 & 4). The exterior of the Villa Termiti consists of
stucco walls and a shingled roof. Wooden panels with a door and two
windows are detachable and centered on each of the four walls. There are
no interior walls, insulation, or fire-blocking. However, the building
does have electrical wiring and a nonfunctional waste-water plastic (ABS)
pipe. The foundation consists of slab and a raised perimeter wall (Fig.
4C).
TheVilla Termiti was designed and constructed by licensed
contractors approved by the Operations Center, Richmond Field Station
(RFS), University of California, Berkeley. Design, construction, building
materials, and safety permits for the Villa Termiti complied with
University of California and Contra Costa County building codes and were
approved by RFS and the Fire Marshall's office, University of California,
Berkeley.
Preparation of Artificially Infested Boards. Kiln-dried, vertical
grain, and clear Douglas fir "'1 X 4s," "2 X 4s," and "4 X 6s" (1.8 X 8.7,
3.8 X 8.7, and 8.5 X 13.7 cm cross-sectional dimensions, respectively) were
cut into 61 cm lengths. On each board, 3 longitudinal cuts were made. The
first cut was a 0.6-cm thick veneer piece cut along the 3.8-cm and 8.5-cm
edge for the 2 X 4s and 4 X 6s, respectively. For 1 X 4s, a 0.3-cm thick
longitudinal cut was made along the 1.8-cm edge. The second cut was also a
0.6-cm thick veneer piece along the 8.7-cm and 13.7-cm edge for the 2 X 4s
and 4 X 6s, respectively. The 1 X 4s had a slightly thinner (0.3 cm) cut
on the 8.7-cm edges. The third longitudinal cut divided the non-veneer
8.7-cm and 13.7-cm side of each board in half (Fig. 5).
Three gallery spaces were routed into each board. The dimensions
of each routed space was 50.8 X 1.9 X 0.6 cm (lwh) for 2 X 4s and 4 X 6s.
Due to the smaller size of the 1 X 4s, the routed spaces were slightly
smaller: 50.8 X 1.6 X 0.9 cm on the 1.8-cm side and 50.8 X 1.9 X 0.3 cm on
the 8.7-cm side. Each gallery had a unique designation (Fig. 5). Gallery
1 was a centrally-routed space just under the veneer piece on the 1.8-,
3.8- or 8.5-cm side of the 1 X 4s, 2 X 4s, 4 X 6s, respectively. Gallery 2
was also a centrally-routed space just under the veneer piece on the 8.7-
or 13.7-cm sides of the boards. Gallery 3 was the routed space centrally
located in the half of the board that did not contain Gallery 2 (Fig. 5).
Since individual treatments may have varying penetration within
wood, the exposure of individual galleries to treatment could be important.
The random and asymmetrical positioning of galleries within boards allowed
us to determine how gallery orientation and wood thickness affected
treatment performance. Prior to installation, boards were held together
with two rubber bands and stored at ambient environmental conditions in the
laboratory away from excessive light. With the exception of boards used in
the untreated group, no individual board was exposed to more than one
treatment or used in more than one test.
Insertion of Insects into Boards. Seventy-five drywood termites
were placed within each board, 25 into each of the three routed galleries.
Once the termites were in place, boards were held together with a 2.5-cm
wide masking tape and individually labeled with a unique identification
number and treatment type.
Placement in the Villa Termiti. Since all treatment methods varied
in application and mode of action, two testing options, A and B, were
offered to vendors participating in this study. Option A consisted of
localized treatment of test boards only behind drywall in the Villa Termiti
wall voids. Option B consisted of a whole-structure treatment, with test
(or infested) boards positioned in the attic, the wall voids of the "living
space" and the subarea.
For Option A, 24 test boards were installed in the Villa Termiti.
The dimensional sizes of boards used were: four 1 X 4s, sixteen 2 X 4s and
four 4 X 6s. Excluding the detachable segments (the four side panels each
with one door and two windows), 24 wall voids (six in each wall) in the
drywall area of the Villa Termiti could be used (Fig. 3). The experimental
design required the use of two wall voids from each of the four sides of
the building (8 wall void spaces total). These 8 wall voids were randomly
selected from the available 24 voids, excluding the 6 wall voids which
contained electrical wiring. Three test boards were placed into each of
the randomly selected wall voids: two 2 X 4s and either one 1 X 4 or one 4
X 6. All test boards were also randomly selected.
Within each void space, there were four possible positions for
board placement, the upper and lower locations for both the right and left
studs (Fig. 3). The upper locations were positioned mid-stud and lower
locations rested on the sill plate. Three of these positions were randomly
chosen for installation of three test boards.
The orientation of galleries in the test boards within a wall void
was also randomized for each board. When viewed straight on, Gallery 1 had
two possible positions: facing the back wall of the void space or rotated
180 degrees and facing the installer. Gallery 2 also had two possible
positions, affixed to the stud or rotated 180 degrees directed away from
the stud. Because Gallery 3 was centrally located within boards, its
orientation was not greatly affected by rotation. Each board was affixed
to the studs with two 0.3-cm diameter drywall screws 5.1-cm long.
For Option B, the same methods employed in Option A were used.
Twenty-four boards were installed in wall voids behind the drywall in the
"living area." In addition, 12 boards were installed in the attic and the
subarea. Thus, a total of 48 boards were installed within the entire
structure. Treatment locations of test boards in the attic were exposed
void spaces located near the mid-line of the gable roof (Fig 3.). Two
possible sampling locations, the right or left void space, were used in
each of the four exterior walls of the attic. After randomly selecting a
void space, three test boards were installed: two 2 X 4s and either one 1 X
4 or one 4 X 6. There were three possible board locations within the void
space: the right or left stud, or against the void's exterior wall.
Individual boards were randomly assigned to a location. Two drywall
screws, described in Option A, were used to affix boards to test locations.
 
In the subarea, test board installation consisted of affixing
boards to the mudsill; overhanging them onto the concrete foundation or
affixing them along the mudsill plate (Fig. 3). There were three test
locations per side of the building: right and left locations on the
concrete wall and the middle of the mudsill plate.
As mentioned above, boards were randomly selected and galleries
randomly oriented prior to installation. For boards overhanging the
concrete, Gallery 1 had two possible orientations, up or rotated down 180
degrees in a vertical plane. Gallery 2, when viewed straight on, either
faced towards or was rotated 180 degrees away from the concrete wall. The
orientation of Gallery 3 remained unchanged through rotation. Boards
installed centrally on the mudsill plate were affixed midway in a
horizontal plane. The orientation for Gallery 1 when viewed straight on,
either faced the installer or was rotated 180 degrees towards the outside
wall. Gallery 2 also had two possible orientations, down towards the
mudsill or rotated 180 degrees facing upwards in a vertical plane. The
orientation of Gallery 3 remained unchanged through rotation. Two drywall
screws, aforementioned, were used to affix boards in place.
Treatment boards were installed in the Villa Termiti approximately
24 h before testing. Untreated boards were left undisturbed in a separate
building approximately 30 m from the Villa Termiti.
Placement of Naturally Infested Boards. The criteria used for
selecting naturally infested boards for the study were: 1) standard
dimensional lumber, 2) 1.8 X 18.4 cm, but not more than 8.5 X 13.7 cm in
cross-section, and 3) acoustical emission readings greater than 10 counts
per min in at least one monitored position within the board (Scheffrahn et
al. 1993). Boards were further stratified into low, medium, and high
acoustic activity. Corresponding acoustic emission readings for stratified
levels were approximately 10, 30, and >40 counts/min as registered by a
hand-held acoustic emission detector (Wood-destroying Insect Detector®,
DowElanco Indianapolis, IN). When possible, an equal number of boards
within each stratum were installed in the three areas of the Villa Termiti.
An additional consideration in the placement of boards was its ability to
fit within selected test positions in the Villa Termiti.
The actual test positions within the Villa Termiti varied between
whole-structure and localized treatment methods. For whole-structure
treatments, boards were installed in the attic, "living space," and
subarea. Three boards were placed in the attic, two in spaces between the
ceiling joists (one on the east side of the building and one on the west
side) and one in an exposed north-facing wall space (Fig. 3). In the
"living space", three boards were installed: one on the east and west
header beneath the drywall and one in a wall void (Fig. 3). For the
subarea, three boards were installed, one on each east and west 4 X 8 (8.3
X 18.4 cm) subfloor support joists and one laid on top of the north-facing
mudsill (Fig. 3).
All test boards were randomly selected and test positions randomly
assigned among boards. Localized treatments, for the most part, used the
same test positions (for exceptions, see Vendor Cooperation section). In
total, nine naturally infested boards with measurable termite activity (>10
acoustical counts/min) were used for each test.
Vendor Cooperation. The authors did not conduct any of the
applications. Licensed commercial vendors were solicited for all
applications in the Villa Termiti. All control methods tested are services
offered by firms licensed by the Structural Pest Control Board of the State
of California. Letters soliciting cooperation were written to the vendors
providing the six methods undergoing tests. Cooperation was voluntary.
We felt treatments should represent standard procedures in the
field as outlined by usage levels and vendor training manuals. Before
treatments, we requested detailed information on procedures to be used by
vendors. We also requested that all vendors use procedures that minimize
test board and structural damage. Photographs and video-recording, if
allowed by vendors, were also used to document exact procedures. All
vendor treatments were conducted separately when vendor schedules and
availability of termites allowed. Treatment effects such as temperature
and gas composition were monitored by the vendor. For a complete review of
operating procedures, safety and limitations, individual vendors should be
contacted for complete training and operating manuals. The following are
the names and normal operating procedures for firms that agreed to
participate in this efficacy study.
Fumigant gases. Sulfuryl fluoride, Vikane® fumigant gas (a
licensed product of DowElanco), was one of the two fumigants used during
the study. This fumigant is colorless, odorless and extremely harmful or
fatal to humans; therefore, it must be handled with extreme caution by
trained and certified personnel. This technique is a whole-structure
treatment. Two firms conducted the work during the testing period: Knight
Fumigation and Ultratech Division, both from San Jose, California. In
addition, the Senior Industry Specialist of DowElanco for northern
California also participated.
Normal fumigation procedure for the treatment of homes involves
sealing the house with vinyl-coated nylon tarpaulins, fans (5 amps) for
circulation of the fumigant, gas dosage calculation, infusion of the
warning agent chloropicrin, and aeration and clearing for occupant
re-entry. Dosage rate (g/m3) is dependent upon many factors: condition of
tarps, soil type, soil and air temperature, and wind conditions.
Monitoring gas concentrations is optional.
Many studies have been conducted which describe, in detail, pre-
and post-treatment preparations (Thoms & Scheffrahn 1994). For safety, all
pets, plants, and occupants must be removed from the premises during
treatment. Normal treatment time is approximately 22 h. After treatment,
the structure is aerated and cleared for re-entry according to mandated
regulatory and industry standards (Anonymous 1993). Since the entire
structure was treated, knowledge of the exact location of boards in the
Villa Termiti was not known to the vendor.
The Villa Termiti was treated three times with sulfuryl fluoride
(November 19, 1993, April 11, 1994, and September 29, 1994). In the first
fumigation 48 artificially infested boards were placed in the structure.
For the second treatment, only naturally infested boards (9 total) were
placed. For the last fumigation, both artificially and naturally infested
boards were included: 36 artificially and 9 naturally infested. All
treatment days were clear and sunny. The temperature, soil readings from
the subarea, was 10, 15.5, and 18.9o C, respectively. Wind speed was less
than 1.4 km/hr, except for the November 1993 treatment, when the wind was
recorded at 3.6 km/hr. The tarp conditions were good or excellent. Soil
type was clay and the seal condition was good. The total calculated volume
treated was 198 m3 (the actual volume of the Villa Termiti is 154 m3). The
extra treated volume included the additional tarp space for eave overhangs
and porches. The amount of sulfuryl fluoride used for each fumigation was
7.4, 2.3, and 2.5 kg. Differences in the amounts of the fumigant used
reflected the varying temperature and wind conditions for each day. A
Fumiguide® was used to calculate exact dosages required for successful
treatment. All treatments were monitored with a fumiscope; readings (ppm)
were taken in the attic, drywall, and subarea. Sulfuryl fluoride gas
levels were monitored at approximately 1 h and 22 h (just prior to tarp
removal) after gas insertion. First hour fumiscope readings for all three
treatments, 39.1 g/m3, 14.4 g/m3 and 13.5 g/m3, were at or above
recommended rates. Twenty-two hour post-fumigation fumiscope readings for
the three treatments was 14.0 g/m3, 8.9 g/m3 and 5.9 g/m3, respectively.
The second fumigant used during the study was CO2-synergized methyl
bromide, employing the MAKR® Fumigation Process. This fumigant is also
colorless, odorless, and extremely harmful or fatal to humans and must be
handled with extreme caution by trained and certified personnel.
Participating vendors included A-1 Fumigation and Farmer Pest Control
(Bellflower, California), Cal Ag (Woodland, California), and Discount
Fumigation (San Jose).
The active ingredient of this gas is methyl bromide. To enhance
the effects of the active ingredient and to minimize aeration time and
toxic gas release into the atmosphere, carbon dioxide is added as a
synergist. This synergized mixture allows a two-thirds reduction of the
normal application rate: 7.3 g/m3 reduced to 2.4 g/m3 (24 oz per 1,000 ft3
reduced to 8 oz per 1,000 ft3). The amount of carbon dioxide used is
approximately 10% of the total cubic volume of structure treated.
The Villa Termiti was fumigated three times with CO-synergized
methyl bromide (September 24, 1993, January 20, 1994, and October 13,
1994). The treatment pattern involving separate dates for treating
artificially and naturally infested boards plus one treatment combing both
types of test boards. The number of treated boards was similar to those
described above for sulfuryl fluoride. The climatic conditions for all
treatment days were sunny and clear with air temperatures of 26.7, 15.6,
and 20.0o C, respectively. Wind speed for all treatments was less than 2.4
km/hr. Trap and seal conditions were excellent for all treatments. The
total calculated volume treated was 178 m3. As with the other fumigant,
vinyl-coated nylon tarpaulins were used to enclose the entire structure.
The amount of methyl bromide used was the same for each treatment (1.4 kg)
because this fumigant has only one dosage rate irrespective of climatic
conditions. The amount of CO2 used was 31.3 kg (approximately 53.7 g/m3).
The time of exposure for all treatments was 22 h. The mixture of methyl
bromide and carbon dioxide was heated to a minimum of 70o C and then
introduced into the structure in less than 1-min through a hose. The gases
were dispersed by two 5-amp fans. A warning agent, chloropicrin, was also
added. The presence of methyl bromide and CO-2 gases was documented by
piercing the tarps and taking an internal air sample using a Draeger® tube.
Tube readings for all treatments exceeded 1,500 ppm MB and 10% for CO2
(both values are according to labeled rates). Since the entire structure
was being treated, knowledge of the exact location of boards was not known
to the vendor.
Heat. The Villa Termiti was treated with heat three times. All
test days (October 29, 1993, January 17, 1994, and December 1, 1994) were
sunny and clear. The coldest initial temperatures in the Villa Termiti
prior to heating for each treatment date were 15o, 12.2o, and 8.9o C: all
were found in the same attic beam. J. H. Steffenson Termite Control
(Campbell, California) in conjunction with Isothermics/Thermal Pest
Eradication® (TPE) (Orange, California) applied the heat treatments. Two
vinyl-coated nylon tarpaulins were needed to enclose the structure prior to
treatment. The tarpaulins had several tears to allow air movement through
the structure (hot air was continuously circulated through the Villa
Termiti). Thermocouples were placed throughout the structure to record
temperature changes (Fig. 6). The number of thermocouples used varied with
each test: 6 thermocouples were used during the first test, 10 for the
second, and 11 for the third. Four convection heaters (each 400,000 BTUs),
powered by propane, were positioned outside and hot air blown inside
through flexible Mylar® ducts.
The objective of heat treatment is to have the temperature of the
coolest thermocouple, normally in a large wooden member or mudsill in the
subarea, reach at least 48.9o C and remains at that temperature for 30 min.
Total treatment time is typically about 6 h for a two-bedroom home (Ebeling
1994). Two 4.3-amp fans were positioned in the structure (subarea and
drywall areas) to insure uniform heat distribution. During normal
operations, air temperatures within living spaces are not to exceed 65.6o C
to minimize any damage to the treated structure or its contents. The Villa
Termiti was vacated during treatment; however, entry into the building was
possible, though uncomfortable, even at these elevated temperatures. Since
it as a whole-structure treatment, knowledge of the exact location of test
boards was not known to the vendor.
Liquid Nitrogen. There was no vendor cooperation for this
treatment method. All liquid nitrogen treatments in the Villa Termiti were
conducted by research personnel from the University of California, Berkeley
and Riverside. Since this is a spot-treatment technique, knowledge of the
location of test boards in the Villa Termiti was made known to personnel
conducting the treatments. Optimal performance of liquid nitrogen requires
an enclosed void space for containment of the liquid and vapor. A 1.3-cm
diameter hole was drilled through the drywall near the top plate of the
void space being treated. Liquid nitrogen (Altair Gases & Equipment, Inc.,
Oakland, California) was then injected from 160 l dewars into the wall
cavity through a 1.2-m flexible woven stainless steel hose. The
temperature inside the dewar may be lower than -195.8o C.
In our tests we attempted to have a constant amount of liquid
nitrogen delivered to each wall void. We calibrated the time it took to
deliver a standard dose. In the first test, we delivered liquid nitrogen
at approximately 30 min @ 1.4 kg/min into each 14.5 X 30.5 X 244 cm wall
void (n = 13 separate wall voids). In the second test, we delivered
slightly less than one-half and the third test one-fourth the rate applied
during the first test (0.9 kg/min for either 15 min or 7 min). A digital
scale (Pennsylvania 66000, 1.2 m2 platform, 2,268 kg capacity) was used
during the first test to determine pre- and post-application weights of
liquid nitrogen. A full-capacity beam scale (Howe Richardson Mechano-Weigh
54 XL) was used to record changes in liquid nitrogen weights for the two
additional tests.
Twelve thin-wire thermocouples were used to monitor temperature
changes within studs and test boards in wall voids of the Villa Termiti.
The ends of the wires were inserted into a scanning thermocouple
thermometer (Cole-Parmer model #92800-00) that was connected to a serial
printer via an RS-232 output. Temperature readings were taken every minute
from 12 different locations in "2 X 6-in" (3.6 X 14.5 cm) studs and test
boards for each corner of the building (6 possible void spaces per corner)
(Fig. 7). Control should be reached when the temperature falls below
-28.9o C for at least 5 min (Tallon 1992); the temperature inside the dewar
may be less than -184.4o C.. Insulation blankets were used to help
maintain the frigid temperatures for the last two tests. For safety, the
oxygen content of the living space was monitored, protective clothing was
worn, and the "buddy system" was used.
Electrocution. Etex, Ltd., Las Vegas, Nevada, manufacturer and
distributor of the Electrogun®, supplied all equipment and personnel for
the treatment. The equipment used is commercially marketed as the
Electrogun®, a device that kills drywood termites by emitting high
frequency electricity (100 kHz), high voltage (90,000), but low current (<1
amp). Whole-house and spot-treatment options are offered by vendors. For
exposed 2 X 4s and smaller pieces of wood, the probe end of the device was
placed against the wood surface. For larger pieces of wood and wood
concealed behind drywall, a "drill-and-pin" method was used.
For the "drill-and-pin" method, small (1.6 mm diameter) holes were
drilled through the drywall and into the wood. Approximately 15.2-cm long
straight copper wires were inserted into the holes and into the termite
galleries. Several consecutive drillings per hole were used to insure that
the electrical current was delivered at various depths within the boards.
For tests in the Villa Termiti, knowledge of the location of test boards
behind drywall was revealed to the vendor.
For optimal performance, thorough coverage of infested wood is
crucial. This method may also be used with other treatments such as
chemical liquids or dusts that can be applied into galleries or onto the
surface of infested timbers. The training information and equipment manual
do not recommend using this method in close proximity to metals, concrete,
excessive moisture, laminated finishes or electronic items. Use in
inaccessible areas is also not recommended (nor is it practical). This
treatment method is available only to licensed professional operators who
lease the equipment and undergo an extensive mandatory training program.
When properly used, this device is reportedly safe for operator and
building occupants; no potentially harmful radiation is emitted.
An additional and different test in the Villa Termiti was requested
by the Etex Corporation because they felt the conditions of the first test
produced results biased against the Electrogun®. They were concerned about
proximity of test boards to metal (wire-mesh in stucco and metal support
bars in detachable walls) and to concrete in the foundation of the subarea.
They felt the performance of the Electrogun® would be improved if tested
against infested boards in locations without interfering metal or concrete.
For the second electrocution test, 18 artificially infested boards
were installed in locations away from sources of metal and concrete. Six
boards each were installed in the attic, "living space," and subarea. In
the attic, boards were randomly placed on one of two 2.7-m long rafters
adjacent to the vertical wall for each side of the building; thus, there
was a total of 8 possible sampling locations (Fig. 3). In the "living
space", each board was installed in a wall void (2.0 X 0.2 X 0.1 m) in the
detachable walls only. There were 16 total wall voids possible for
selection; four for each side of the building (Fig. 3). Except for the
metal window frames and nails, the detachable walls do not contain
significant amounts of metal. For the subarea, floor joists were used for
board installation; no boards were placed on or near the sill plates or
foundation. Six joists were randomly chosen from a possible total of 46.
The preparation, fastening, randomization, and orientation of test boards
were similar to the earlier test. The vendor was asked to treat the boards
as before. Fifteen controls, or untreated boards, were kept in the same
nearby building (see above).
Naturally infested boards were similarly installed. In total, nine
naturally infested boards were installed in the Villa Termiti for this
second test: three boards each in the attic, living space, and subarea. In
the attic, two ceiling bays and one rafter were used. In the living space,
three void spaces in the detachable walls (different from those containing
artificial boards) were used. In the subarea, three floor joists
(different from those containing artificial boards) were used. The
preparation, fastening, randomization, and orientation of boards were also
similar to the earlier test.
Microwaves. The Termite Inspector (Mission Hills, California)
supplied the equipment and personnel to conduct all tests in the Villa
Termiti. Treatment procedures included treating infested wood with a
700-watt (more powerful devices are available from other vendors). The 2.4
GHz frequency oscillation causes vibration of water molecules within
termites, which produces lethal temperatures above 48.9o C. This
spot-application method treats a section of wood approximately 10.2 X 30.5
cm when operating. For safety, the device is operated remotely, and all
persons are kept a minimum of 9.1 m from the apparatus while it is
operating. Protective blankets are used, if necessary, for added safety.
Under field conditions, the vendor uses visual, acoustical, or fiber
optical devices to locate infestations. For tests in the Villa Termiti,
the location of test boards was disclosed to the vendor. Once boards were
identified for treatment, the microwave instrument was positioned over the
treatment spot, technicians left the room, and the device was operated by
remote. Treatment time was approximately 8 min per spot (10.2 X 30.5 cm;
311 cm2). A hand-held microwave detection device (Model HI-1801, Holaday
Industries, Inc.) was used to selectively monitor emissions from the
opposite side of the wall during treatment.
Assessment of Treatment Efficacy. The day following treatment, all
artificially infested boards (24 to 48 total) were removed from the Villa
Termiti and stored in the laboratory until they were opened for 3-d
post-treatment assessment of mortality. Live and dead termites were
counted and removed from each gallery in every board. Live termites from
each gallery were placed in a separate tongue depressor holding chamber
(Fig. 1). each individual holding chamber was stored in a clear, 15.5-cm
diameter plastic container with a lid and placed in an incubator stored in
a glass greenhouse. The percent mortality of these termites was determined
at 4-wk post-treatment. Percent mortality was defined to include both
carcasses and those missing. Since termites are cannibalistic and eat
their dead and injured, we used this measure of mortality. Termites
smashed by handling and recording miscounts (gained numbers of termites)
were excluded from the analysis. Dead insects were stored in clear 23 X 23
X 20 mm plastic boxes with lids for future verification. Untreated boards
("controls") containing 25 termites in each of three galleries were
assessed in the same manner as treated boards.
After each test, the nine naturally infested boards were removed
from the Villa Termiti and stored in the same glass greenhouse as the
artificial boards. Boards were acoustically monitored at 1-, 2-, and 4-wk
post-treatment (these data will be reported separately). At 4-wk
post-treatment, the boards were cut into small lengths (approximately
10-cm) and carefully dissected. The sections of board were split with a
hammer and wood chisel until it was clear there were no more termites in
every gallery and chamber in the board. Live and dead termites were
counted and sorted by caste (alate, soldier, and nymph/pseudergate) and
percent mortality was calculated based on these data for each board. All
insects, dead and alive, were stored in labeled, clear 23 X 23 X 20 mm
plastic boxes with lids for future verification.
For each artificially infested board, the percent mortality was
calculated by combining the counts of live and dead termites for all three
galleries. Thus, the experimental unit for this portion of the test was
the board. In addition, separate records were kept for each gallery so
that we could explain anomalies in mortality as a function of the unique
placement of a particular gallery within the wall voids or near the
foundation. Occasionally the termites burrowed through the sawed portion
of the board and moved from one gallery to another. An additional
complication was that as a result of the "drill-and-pin" method of applying
the Electrogun®, holes were left in the boards, allowing termites to
escape. If the total number of termites remaining in an individual
artificially infested board was less than 48, the data for that particular
board was discarded for analysis. Galleries not containing any termites
were also discarded from further analysis. Since we had little control
over the number of termites in a naturally infested board, we calculated
percent mortality using all of the live and dead termites removed from the
board.
For each treatment, the weighted mean response ( ) and the standard
error of the weighted mean ( ) for artificially infested and naturally
infested boards were determined using the following formulae:
 
where
= the mean mortality for a board in the th location in the building and
for the th size board.
= the standard error associated with a particular mortality ( ).
= the proportion of lumber of dimension placed in test within location .
= location in the building: (1) attic, (2) living area, and (3) subarea.
= dimension of lumber: (1) 1 X 4 or 1 X 8; (2) 2 X 4; and (3) 4 X 4, 4 X
6, 4 X 8, or 4 X 12.
 
The relative proportion of wood, 1 X 4s, 2 X 4s, and 4 X 6s, for
each location within the Villa Termiti was calculated using the following
formula and was equal to one.
 
 
 
In calculating a weighted percent mortality for a building, we
assumed equal rates of infestation for all three areas of a home. For a
given locality this may not be true. For example in Riverside and Palm
Springs, California, or Phoenix, Arizona, infestations of drywood termites
would likely be restricted to the subarea or window and door frames.
Infestations are not common in the attic area in these desert locations.
In contrast, drywood termite infestations in coastal California are common
in the attic, walls, and even in the subarea.
We assumed no density dependence in efficacy of any of the
treatments. For the artificially infested boards, this is not a problem
since we kept density constant. For the naturally infested boards we were
not able to regulate density of drywood termite populations. We did
estimate relative densities of the populations in each board before each
test using acoustic emission counts (Lewis et al. 1991). We then
stratified the boards used for a particular test into low, medium, and high
populations. We positioned these boards, stratified on the basis of
apparent termite density, throughout the Villa Termiti by placing an equal
number of each category in each location/level.
Treatments were not statistically compared to one another. The
efficacy of each treatment was tested against a set of standards. Our null
hypothesis was that each treatment is ineffective, i.e. the treatment
resulted in mortality less than or equal to 90 percent (Ho: p < 0.90). The
alternative hypothesis was that each treatment resulted in a mortality
greater than 90, 95 or 99 percent (Ha: p > 0.90). The equation used to
test the significance of the hypothesis is:
 
 
where
m = level of mortality desired or required.
 
All statistical tests of efficacy were determined by a one-tailed
t-test to determine whether the mean response exceeded the standard level
of mortality (Steel and Torrie 1960). Statistical significance was tested
at the a = 0.05 level with degrees of freedom equal to the number of boards
minus 1. The number of boards, or replicates, varied among treatment
methods. For spot-treatments, a replicate was an individual board. For
whole-structure treatments, one application of the method to the entire
structure would ideally be considered a replicate. However, for this
study, each board for all treatment methods was considered a replicate.
Summary statistics for mortality levels among location, board
dimension, and gallery designation were derived with the MEAN procedure
(PROC MEAN, SAS Institute 1994). Means for termite mortality levels among
Villa Termiti locations, board dimension, and gallery designation were
analyzed for significant differences using the Ryan-Einot-Gabriel-Welsch Q
multiple range test (PROC GLM, SAS Institute 1994). Untreated board
replicates for each treatment method were pooled and mortality levels
analyzed by board dimensional size and gallery designation using
Ryan-Einot-Gabriel-Welsch Q multiple range test (PROC GLM, SAS Institute
1994). All artificially infested boards were visually inspected after
treatment for signs of damage: drilled holes and burn marks. Differences
in the number of damaged boards among treatment methods were analyzed using
the Ryan-Einot-Gabriel-Welsch Q multiple range test (PROC GLM, SAS
Institute 1994). Differences in treatment time and number of drilled holes
between the two electrocution tests were analyzed using paired t-tests
(PROC TTEST, SAS Institute 1994).
Results and Discussion
 
Fumigant Gases
Sulfuryl fluoride. Termite mortality for all treated
artificially infested boards was 100% (Table 1). The overall mean
mortality value for the entire structure was significantly greater than the
90, 95, and 99% levels of acceptance (Table 2). Therefore, the efficacy of
this treatment significantly exceeds the 99% mortality level. These data
agree with previously reported laboratory studies in sulfuryl fluoride
efficacy (Su & Scheffrahn 1986, Osbrink et al. 1987, Thoms & Scheffrahn
1994). Control mortality at 3-d was low (< 5%), suggesting high
survivorship in test boards prior to treatment. However, mortality after
4-wk increased more than 5-fold, suggesting increased natural and handling
mortality post-treatment. Differences in mortality among board dimensions
for controls were nonsignificant (F test, P > 0.05) at 3-d and 4-wk
post-treatment.
The overall mean mortality value for naturally infested boards was
also significantly greater than the 99% (Tables 2 & 3). Only a solitary
soldier survived. Although a single survivor among thousands is
insignificant, this finding is scientifically curious. Lethal sulfuryl
fluoride dosages for soldier castes for a number of termite species have
been previously reported (Osbrink et al. 1987). However, values for I.
minor were not included. Perhaps future studies may include the
determination of the lethal dose for the soldier caste for I. minor.
Except for warping, no visual signs of damage were noted for test
boards treated with sulfuryl fluoride (Table 4). However, all test boards
experienced some signs of wood distortion, especially the thin veneer
pieces. Desorption and residual studies for sulfuryl fluoride report its
safety for many household commodities if properly used and aerated after
treatment (Kenaga 1957, Osbrink et al. 1988, Scheffrahn et al. 1987, 1989a,
1989b).
CO2-synergized methyl bromide. Termite mortality for all
artificially infested boards was 100% (Table 5). The overall mean
mortality value was significantly greater than the 99% level of efficacy
(Table 2). Control board mortality at 3-d was moderate and ranged from 8.0
to 16.2% (Table 5). Natural and handling mortalities for controls
increased at least 2-fold 4-wk post-treatment. The differences in
mortality among board dimension for control boards for 3-d and 4-wk
post-treatment was nonsignificant (P > 0.05). The mortality level for
naturally infested boards was also significantly above the 99% level
(Tables 2 & 6). However, 30 survivors were found in one 2 by 4 in the
subarea. None of the survivors were reproductives. Therefore, we conclude
that this synergized, reduced application rate of methyl bromide
significantly exceeds the 99% mortality level.
Dose mortality curves for drywood termites (I. synderi) using
CO2-synergized methyl bromide have been reported (Scheffrahn et al. 1995).
Scheffrahn & Su (1992) have also reported that label rates for methyl
bromide were excessive by as much as 4-fold. Both papers support claims of
high levels of efficacy for reduced methyl bromide dosages. Results from
this study support these earlier findings and suggest that care must be
taken in calculating dosage, as well as placement and number of fans when
using reduced methyl bromide methods, especially in subareas.
Visual signs of damage to test boards was minimal for this
treatment and restricted to warping of some 2 by 4 and 4 by 6 test boards
(Table 4). However, other studies have shown that some household foods, if
not enclosed in protective nylon bags, can serve as sorptive matrices for
this fumigant (Scheffrahn et al. 1990, 1992). (Rudi, can your "Indoor
airborne residues of methyl bromide and sulfuryl fluoride following
aeration of fumigated houses" paper be cited?). Methyl bromide has also
been reported to be an ozone depletor (Cicerone 1987). Registration of
methyl bromide is scheduled for phase-out by the U. S. Environmental
Protection Agency (EPA) by the end of the decade (Kramer 1992).
 
Nonchemical Methods
 
Excessive heat. Termite mortality in artificially infested boards
was 100 percent except in the subarea (Table 7). Mortality levels were
significantly greater than the 90 percent efficacy levels after 3-d and
significantly greater than the 90 and 95 percent levels after 4-wk (Table
2). The subarea was the only location that mortality levels did not reach
100 percent. Mortality values for the subarea at 3-d and 4-wk
post-treatment was 85.8 and 91.1 percent. Both values were significantly
different from mortality values for the attic and "living space" (F = 17.6;
df = 2, 80; P < 0.0001; F = 11.4; df = 2, 83; P < 0.0001). It was
initially thought that the early removal of test boards during the first
test, immediately after treatment to protect termites from predation by
Argentine ants, Linepithema humile (formerly Iridomyremex humilus), may
have interfered with test results. However, results from the second test,
when boards were not removed early, revealed a similar pattern of
survivorship in the subarea.
Analysis of data for the subarea revealed an uneven distribution of
mortality for artificially infested test boards. The size of the
artificially infested boards did not have a significant impact on mortality
levels achieved at 3-d or 4-wk post-treatment (P > 0.05)(Table 7).
However, the data for individual galleries within boards revealed that
survivorship only occurred in Gallery 2; locations affixed against the
foundation wall (Fig. 4). At 3-d post-treatment the mortality value for
Gallery 2 (78.2 percent) was significantly lower than Gallery 1 (94.5
percent) and Gallery 3 (92.2 percent)(F = 7.1; df = 2, 208; P < 0.0011).
Similarly at 4-wk post-treatment the mortality value for Gallery 2 (80.7
percent) was significantly lower than Gallery 1 (95.3 percent) and Gallery
3 (92.8 percent) (F = 6.4; df = 2, 208; P < 0.002). Data from
thermocouples revealed that all probes reached the 50o C lethal temperature
for at least 1 h (Table 8); however there were still survivors. It is not
known how much more additional time would have been required to achieve 100
percent mortality for test boards in the subarea. Perhaps future studies
can focus more on thermocouple position and number in subareas and sill
plates on slabs.
Termite mortality for controls was initially high for 1 by 4s
(Table 7). Four-week control mortality values among boards, albeit were
large and ranged from 24.2 percent in 2 by 4s to 43.5 in 1 by 4s, were not
significant (P > 0.05). These results, probably due to overheating of some
boards while in transit to the laboratory from the Villa Termiti, suggest
termite robustness was less for the heat treatment than for other methods
tested.
Mortality results for naturally infested boards was 100 percent for
all test locations in the Villa Termiti (Table 9). Mean mortality was
statistically significant at the 90, 95, and 99 percent levels of efficacy
(Table 2). From the results with artificially infested and naturally
infested boards we conclude that excessive heat, applied as described,
results in a mortality level that significantly exceeds 95 percent and
perhaps as high as 99 percent.
There were a few visual signs of damage, minor warping for some 2
by 4 and 4 by 6 test boards (Table 4). Other changes noted in the Villa
Termiti included sticking of doors (reversible), fluorescent lights
going-out (reversible) and warping of a non-functional ABS plastic
waste-water pipe (non-reversible). However, under normal field conditions,
a low volume of cold running water is left on to prevent the warping of
plastic pipes. Minor structural damage from heat treatment, as well as
pre-treatment preparations to minimize damage to household items, have been
previously reported (Forbes & Ebeling 1987, Ebeling 1994).
Excessive Cold. Our assessment of the effectiveness of
spot-treatments with liquid nitrogen was mixed and highly influenced by
dosage and thermocouple placement. At the highest dosage tested, 30 min
@1.2 kg/min, both 3-d and 4-wk mortality of drywood termites in
artificially infested boards was 100 percent (Table 10). At this dosage,
mortality was significant for all efficacy levels tested (Table 2).
Similarly, the 3-d and 4-wk mortality levels for the 15-min @ 0.9 kg/min
dosage was statistically significant at the 90 and 95 percent efficacy
levels (Tables 2 and 10). However, for the lowest dosage of 7-min @ 0.9
kg/min, the 3-d mortality value, 84.4 percent, was much less efficacious
(Table 10); this value was not statistically different at the 90 percent
level (Table 2). The 4-wk mortality value, 87 percent, although slightly
higher, was still not statistically significant at the 90 percent level of
efficacy (Tables 2 and 10). Control mortality levels at 3-d post-treatment
were less than 5 percent, suggesting termites were robust prior to
treatments. However, 4-wk mortality increased approximately 5-fold for
controls indicating considerable natural and handling mortality. There
were no significant difference in mortality levels among board dimensional
size or gallery designation for control boards (P > 0.05). Rust et al.
(1995) contain data tables that report the minimum dosage rate required to
achieve 100 percent control with liquid nitrogen was at least 21 min @0.9
kg/min in an uninsulated 2.4 by 2.0 m (8-ft by 6.6-ft) artificial wall used
in their tests. Their results are clearly in agreement with those reported
in Table 10. Lower dosage rates and application times are not likely to
achieve the minimal lethal temperature.
Lumber dimensions appear to affect termite mortality, at least for
the two lower dosages of liquid nitrogen tested. The 4 by 6 boards
suffered less mortality than the other board dimensional sizes when treated
with the 15-min and 7-min @ 0.9 kg/min dosages (Table 10). The termites in
the 1 by 4 boards always experienced the greatest level of mortality in
these two lower dosages. However these differences were not statistically
significant (P > 0.05). Wood is a poor thermal conductor (Forest Products
Laboratory 1987). Results from the current suggest that wood, if
relatively sound, may provide termites with insulation and protection from
the lethal effects of excessive cold.
Considerable variance in temperature was recorded in wall voids for
all liquid nitrogen dosages tested (Tables 11, 12, and 13). Treated boards
containing live termites for the 15-min and 7-min @ 0.9 kg/min dosage rates
appeared to be associated with thermocouples failing to achieve lethal
temperatures (Tables 12 and 13). However, for the 7-min @ 0.9 kg/min dose,
three artificially infested boards (two 2 by 4s and one 4 by 6) contained
termite survivors. These boards were located in treated wall voids
containing thermocouples that reported minimum lethal temperatures (S3 and
W2, Table 13). These results suggest that higher dosage rates and
thermocouple placement are critical for achieving high levels of efficacy.
Naturally infested boards revealed a similar pattern to that of
artificial boards; decreasing levels of mortality with decreasing dosage
rates (Table 14). For the 30-min @1.4 kg/min rate, termite mortality was
100 percent. The overall mean mortality value was significantly greater
than 90, 95 and 99 percent (Table 2). Similarly, at the 15-min @ 0.9
kg/min dosage, mortality was significantly greater than all efficacy levels
tested (Table 2). However, the 7-min @ 0.9 kg/min dosage rate achieved an
overall mortality level of 74.3 percent (Table 14). This mortality rate
was not significantly greater than the 90 percent level of efficacy (Table
2) and live alates were found among the survivors for the 15-min and 7-min
@ 0.9 kg/min rates. Similar to the results with artificially infested
boards, the existence of termite survivors was associated with failure of
achieving minimum lethal temperatures. Three termite survivors were found
in a naturally infested 2 by 4 at the 15-min @ 0.9 kg/min dose rate albeit
lethal minimum temperatures were achieved for top and bottom positioned
thermocouples (N3, Table 12).
Because we were not able to obtain information on application rates
from a vendor who applies liquid nitrogen, we had to assess different
dosage rates to determine a minimum application rate that was efficacious.
The wall voids in the Villa Termiti are approximately 14 by 28 by 224 cm
with an internal volume of 87,808 cm2 (87.8 l). Clearly, the 7-min @ 0.9
kg/min (63 kg of liquid nitrogen) dosage is not an effective treatment for
any reasonable level of mortality that would be desired. We feel that the
15-min @ 0.9 kg/min (13.5 kg of liquid nitrogen) application rate is the
absolute minimum to achieve a reasonable level of mortality (> 95 percent).
This application rate places liquid nitrogen into the wall void at a
concentration of 154 g of liquid nitrogen/liter. Application rates
exceeding this level are more likely to provide mortality levels in excess
of 99 percent.
Visual damage to boards from liquid nitrogen treatments was minimal
(Table 4). Frost formation during treatment can be considerable and may
cause damage to some wall coverings. However, the possibility exists for
microscopic wood damage that may result in structural failure during stress
(e.g., earthquakes). Future studies are needed to explore microscopic
damage and the possible loss of wood strength for varying dosage levels of
liquid nitrogen. With this treatment, repair of drilled insertion holes is
required.
Electrocution. Efficacy of electrocution treatments from the first
test were below the minimum level of acceptance for artificially infested
boards. Drywood termite mortality levels at 3-d post-treatment in
artificially infested boards were well below 50 percent in the attic and
subarea (Table 15). Only in the drywall area did mortality levels reach 50
percent or higher. The overall mortality value for the entire structure
was 43.8 percent 3-d post-treatment (Table 2). Four weeks post-treatment,
mortality levels increased to 81.2 percent. However, this efficacy value
was still significantly below the minimum 90 percent level of acceptance
(Table 2). For treated locations within the Villa Termiti, the attic,
drywall, and subarea locations had mortality values of 25.5, 54.2, and 46.3
respectively. At 3-d post-treatment, mortality values for the drywall and
subarea locations were significantly greater than the attic (F = 9.6; df =
2, 142; P < 0.0001). The were no significant difference in mortality
values among board dimensional sizes (P > 0.05). However, gallery
differences in percent mortality within boards were considerable; 61.8
percent for gallery 1, 26. 8 percent for gallery 2, and 46.0 percent for
gallery 3. All gallery mortality levels were significantly different from
each other (F = 13.9; df = 2 142; P < 0.0001). Similarly, 4-wk
post-treatment, mortality values among treated areas within the Villa
Termiti and gallery locations within boards were still significantly
different (F = 31.3; df = 3, 187; P < 0.0001; F = 13.6; df = 2, 187; P <
0.0001). Variable results while using electrocution ranging from 3 - 100
percent mortality for boards artificially infested with drywood termites
have previously been reported (Ebeling 1983). Mortality results for
naturally infested boards in the first test resulted in a similar pattern
to artificially infested boards of low mortality (Table 16); 8 of 9 boards
contained termite survivors while 2 of these 8 boards had several hundred
survivors. The overall mortality level, 88.6 percent at 4-wk
post-treatment, did not significantly exceed the 90 percent level of
efficacy (Table 2).
There are several reasons that may explain the poor performance of
electrocution during the first test. First, penetration of electric
current into wood is limited. Ebeling (1983) reported that the surface
application of electricity is restricted to only 1.3 cm deep into wood.
However, for the current study, the size of boards containing termites was
not a factor; there were no significant differences in mortality among
board dimensional sizes (P > 0.05)). The depth of galleries containing
termites also appeared to be unrelated to termite mortality since the
deepest gallery, Gallery 3, had a higher mortality percentage than Gallery
2, the shallowest. In fact, the drywall locations, sites on non-exposed
wood had significantly higher levels of mortality than the exposed test
boards in the attic (3-d and 4-wk results, Table 15).
A second possible limiting factor for electrocution treatments is
delayed mortality (Ebeling 1983). However, the increased mortality
observed at 4-wk post-treatment, as high as 4-fold, was probably not due to
the effects of electrocution. Levels of mortality in the controls were
also higher, as high as 60-fold, and suggest that increased mortality seen
in boards treated by electrocution at 4-wk post-treatment was due to
natural mortality and handling (Table 15).
The results from the first test of both artificially and naturally
infested boards were challenged by the vendor conducting the electrocution
techniques as being bias against the electrocution techniques because
standard operating procedures were not strictly adhered to. The reasons
given for this poor performance (the presence of test boards next to stucco
walls, wire supports in stucco, and concrete in the subarea) suggest that
electric current was drawn away from the treated areas and thus away from
the targeted termites. A second test, smaller in scope and with test
boards positioned in more favorable, "wooden" locations was conducted to
remedy these concerns.
The results from a second test of electrocution of boards in
locations away from metal and concrete were much improved. Three-day
assessment of artificially infested boards resulted in mortality levels
that were not significantly above the 90 percent level (Tables 2, 16).
Mortality levels at 4-wk post-treatment (98.5 percent) significantly
exceeded the 90 and 95 percent levels of efficacy (Table 2).
Mortality of naturally infested boards was also higher than in the
first test (Table 16). Only 5 boards contained survivors, as compared to 8
in the first test. However, one board contained over 100 survivors. The
mortality level in naturally infested boards for the entire structure was
95.1 percent and significantly exceeded only the 90 percent level of
efficacy (Table 2).
The improved performance of electrocution in the second test
requires some discussion. Ebeling (1983) claimed the mortality effects of
electrocution are heightened when test boards were placed on a metal table.
The results of the current study suggest an opposite interpretation: metal
impedes the effects of electrocution. However, this finding is confounded
by 2 factors: significantly more time was spent treating test boards in the
second test as opposed to the first (7.5 min vs. 20.0 min, t = 5.9; df = 1,
38; p < 0.0001) and significantly more holes were drilled per board during
the second test (6.5 versus 12.0, t = 3.3; df = 1, 61; p < 0.0015). Range
in treatment time for each board varied from 5 min to 1.75 h. Since most
of the test boards were treated with the drill-and-pin technique (59 of 66
for artificially infested boards and 14 of 18 naturally infested boards),
statements cannot be made about passing the probe end emitting electricity
over boards when treating. Future studies are needed to determine the
effects of treatment exposure time and insertion of metal pins upon termite
mortality.
We conclude that the efficacy of this treatment appears to be
technique-driven. Clearly, electrocution causes mortality in termites.
However, to achieve reasonable levels of mortality the operator should use
the drill-and-pin technique and spend as much time as possible treating an
infested area. This control method, more than any of the other tests in
this study, requires precise information as to the extent and location of
the drywood termite infestation. Without accurate delimiting of the
infestation, efficacy will surely drop to unacceptable levels.
Damage to test boards using electrocution was considerable (Table
4). Eighty percent (53 of 66) of artificially infested boards and
seventy-eight percent (14 of 18) of naturally infested boards, revealed
visual signs of damage. Most signs of damage were drill holes from
administration of the drill-and-pin technique. However, 29 boards revealed
minor burn marks within test boards. Ebeling (1983) reported wood is a
poor insulator and could be carbonized or destroyed as a current seeks a
path to ground during treatment. Future studies including varying time
exposures and currents are necessary to more fully understand the effects
of electrocution on wood strength and appearance, especially for wood in
concealed locations.
Microwaves. Considerable variability in mortality was observed
in artificially infested boards treated by microwaves. Mortality at 3-d
and 4-wk post-treatment did not significantly exceed the 90 percent level
of efficacy (Tables 2 and 18). There were no significant differences in
termite mortality between the attic and "living space" at 3-d or 4-wk
post-treatment (P > 0.05). However there was considerable variance (SD) in
mortality especially among 1 by 4s in the attic and "living space" and 2 by
4s in the "living space." 1 by 4 test boards in the "living space" had
lower mortality values than 2 by 4s and 4 by 6s but this difference was not
significant (P > 0.05). Control mortality values were low (< 5 percent at
3-d post-treatment) and suggest high survivorship in test boards prior to
treatment. However, a 5-fold increase 4-wk post-treatment suggest elevated
mortality due to handling and natural causes (Table 18). Mortality values
among control boards of various size dimensions were not significant (P >
0.05).
The overall mortality value for naturally infested boards was 97.4
percent (Table 19). This mortality value significantly exceeded the 90 and
95 percent levels of efficacy. Forty-five survivors were found among three
boards containing live termites, none of which were alates.
The results of this study were mixed. Without a doubt, microwave
energy applied to infested wood will kill termites. Mortality of termites
in artificially infested boards approached 90 percent but did not exceed 90
percent with statistical certainty. Mortality of termites in naturally
infested boards was good, statistically exceeding the 95 percent level of
control. As with electrocution, definition of the extent of an
infestation, location of the infested wood, and access to the infested wood
are all critical to achievement of the desired level of mortality with
microwave treatments.
Visual signs of damage were noted for some artificial boards (Table
4). Minor warping of test boards was noted and 6 boards were burned, 2
severely (Table 4). For shielded ovens, there is no uniform heating of
materials in the microwave field due to internal oven reflection and
non-uniform absorption of energy (Locatelli & Traversa 1989). From the
current study, it appears that unshielded microwave devices also may
produce non-uniform heating of boards during treatment for termites.
Monitored microwave emissions were not detectable at the 9.1 m required
safe operating distance and were less than 6 mw/cm2 within 0.3 m of a
treated wall.
This is the first published report on the efficacy and safety of
microwaves for the control of drywood termites. This method of nonchemical
control appears to have promise as an effective spot-treatment technique.
However, more information is needed to determine the correct time necessary
to achieve the desired level of control and wood penetration of microwave
energy for varying levels of wattage output. Monitoring of temperature
changes in building materials during treatment could improve efficacy and
promoted increased safety.
Handling and Controls. During the course of investigation, over
86,000 termites were handled. I. minor, like most termites, are fragile
and excessive handling can result in mortality (REFERENCE?). Sources of
handling mortality include smashed individuals and miscounts (Table 20).
Less than 3 percent of all handled insects for treated artificially
infested boards resulted in non-treatment mortality. The greatest source
of handling mortality was from the smashing of individuals between the
veneer sections of boards. Boards placed in the subarea had significantly
more smashed individuals than the attic and drywall areas (F = test?).
Dragging boards in the subarea prior to installation was probably
responsible for the increased levels of smashed individuals as compared to
control boards. More secure closing of test boards might/could have
reduced this source of handling mortality.
Control boards were placed into the Villa Termiti, but removed
before treatment -- check hard copy. Handling mortality of control boards
was also low; less than 7 percent. For control boards, the greatest source
of handling error was missing individuals; these individuals could be the
result of miscounts or cannibalism. For this current study, missing
individuals were more likely to have resulted from cannibalism because
"gained individual" counts, another form of miscounts, were very low for
both treated and control boards. Since handling mortality was minimal, we
conclude that any post-treatment mortality observed was primarily due to
natural or delayed treatment effects, not experimental error.
For naturally infested boards, the accuracy of the detection
equipment used (Wood-Destroying Insect Detector®) was also investigated.
Over 94 percent (102/108) of treated, naturally infested boards contained
termites when dissected. Boards determined to contain active termite
infestations by acoustic emissions and used as controls did, in fact,
contain termites 100 percent of the time upon dissection (Table 21). False
positives occurred 17 percent of the time (3/18) (Table 21). Three of
these false positives were boards that contained 1, 2, or 18 live
individuals. Our work substantiates earlier claims on the potential use of
the Wood-Destroying Insect Detector® by identifying pieces of structural
lumber as actively infested by I. minor (Scheffrahn et al. 1993). However,
the authors reported a poor relationship between acoustical counts and
termite population density for detection experiments conducted with I.
synderi (Scheffrahn et al. 1993). From observations of the 140 naturally
infested boards included in this report, a significant correlation (r2 >
0.5) between detection level and termite number was found (Lewis & Haverty,
unpublished data). More technologically advanced acoustic emission devices
may be developed which could improve the predictability of termite
population density in naturally infested wood.
Conclusions
 
Over 90 percent of all test insects for both artificially and
naturally infested boards were killed during treatment. This high
mortality level attests to the fact that all techniques "work" to some
degree. Prior to testing in the Villa Termiti, all vendors claimed high or
total elimination of drywood termite infestations within homes. Test
results from whole-structure treatments for artificially and naturally
infested boards reveal that at the very highest standard of efficacy, 99
percent, only the fumigant gases demonstrated near 100 percent elimination.
Whole-structure treatment with heat was, however, very similar in efficacy
level to the fumigant gases. Perhaps improved application techniques or
the addition of synergists could elevate heat to higher levels of efficacy.
For localized treatments, 30-min @ 1.4 kg/min and 15-min @ 0.9
kg/min dosages of liquid nitrogen and long application times using
drill-and-pin techniques with electrocution were significantly efficacious
at the 95 percent level. Microwave treatments (750 W) were deemed
efficacious at the 95 percent level, but only when treating naturally
infested boards. Not efficacious, even at the 90 percent level, were
electrocution without drill-and-pin techniques and 8-min @ 0.9 kg/min
treatments of liquid nitrogen.
In general, monitored treatments fared better than non-monitored
treatments. Although damage may occur from fumigation, heat or microwaves,
it is a certainty for liquid nitrogen (repairs to drilled holes in wall
voids) and electrocution (drill-and-pin applications). The development of
improved termite monitoring devices in situ could also improve the
performance of all spot-treatment techniques. Field efficacy rates for all
available drywood termite control methods, including drill-and-treat with
chemicals, mode-of action, and damage to structures especially at the
micro-level are potential and important areas for future research. In
addition, standardization of acceptance levels of efficacy and damage to
structures and stricter oversight for advertising claims will be needed.
Acknowledgments
 
The authors wish to thank the many agencies, private firms,
institutions, and individuals that contributed to the study. These include
the Structural Pest Control Board of California for major funding, the Pest
Control Operators of California, Inc., and dozens of private pest control
firms for donations of money for construction of the Villa Termiti. A
special thanks to Director Frank Beall and staff of the Forest Products
Laboratory for accommodating property for the study. We would also like to
thank the tireless efforts of the laboratory crew: Steve Suoja, Edward
Dunbar, Calvin Fouche, Michelle Gaither, Lori Nelson, Carnet Williams, Gail
Getty, Peter Haverty, Ariel Power, Miwako Takano, Miguel Fernandez,
Salvador Rubio-Garcia, Daniel Gaither, Kirsten Copren, Lila Greb, Rebecca
Pollard, Ken Wong, Aikane Lewis, Gail Atilano, Mechelle Cochran, and Thien
Nguyen. Dr. Michael Rust and Eileen Paine from the University of
California at Riverside provided liquid nitrogen treatments. Sources of
drywood termites and naturally infested boards were provided by Ken
Abrahamian, Dr. Thomas Atkinson and dozens of pest control operators
throughout California. Statistical consultations were provided by Dr.
James Baldwin, Pacific Southwest Station, USDA Forest Service, Albany, CA.
The Wood-Destroying Insect Detector® was provided by DowElanco. Reviewers
of earlier drafts of this manuscript include: Dr. Ian Carmichael,
University of California, Berkeley; Dr. Richard Kramer, National Pest
Control Association, Dunn Loring, VA; Dr. Rudolf Scheffrahn, University of
Florida, Ft. Lauderdale, FL; Dr. James Baldwin, Pacific Southwest Station,
USDA Forest Service, Albany, CA. and Dr. Barbara Thorne, University of
Maryland, College Park, MD. Their comments and suggestions have greatly
improved the presentation and interpretation of this study. Lastly, the
authors wish to thank all participating vendors who, very graciously and at
their expense, conducted all treatments.
 
References Cited
Anonymous. 1990. California Statistical Abstract. California Department
of Finance, Sacramento. 204 p.
Anonymous. 1993. Manual for Vikane gas fumigant. DowElanco,
Indianapolis, Indiana.
Atkinson, T. H. 1994. Long Term Effects of Temperature and Humidity on
Survival, Differentiation, and Reproduction of Drywood Termites:
Preliminary Summary (unpublished report).
Atkinson, T. H. 1995. Efficacy of chemical and nonchemical treatments
against drywood termites and incidence of termite damage in southern
California (unpublished report).
Becker, G. 1967. Die Temperatur-Abhangigkeit der Frasstatigkeit einger
Termitenarten. Z. ange. Entomol. 60: 97-123.
Bennett, G. W., E. S. Runstrum & J. A. Wieland. 1983. Pesticide use in
homes. Bull. Entomol. Soc. Am. 29: 31-38.
Bess, H. A. & A. K. Ota. 1960. Fumigation of buildings to control the
dry-wood termite, Cryptotermes brevis. J. Econ. Entomol. 53(4): 503-510.
Bowler, K. 1981. Heat death and cellular heat injury. J. Thermal Biology
6: 171-178.
Brier, A. N. 1987. California homes: characteristics of decay and insect
attack. Wood Building Research Center Publication, Forest Products
Laboratory, University of California, Berkeley. 60 p.
Brier, A. N., W. A. Dost & W. W. Wilcox. 1988. Characteristics of decay
and insect attack in California homes. Calif. Agric. 42(5): 21-22.
Cicerone, R. J. 1987. Changes in stratospheric ozone. Science 237: 35-42.
Collins, M. S., M. I. Haverty, J. P. La Fage & W. L. Nutting. 1973.
High-temperature tolerance in two species of subterranean termites from the
Sonoran Desert in Arizona. Environ. Entomol. 2(6): 1122-1123.
Crocker, R. L., D. L. Morgan, & M. T. Longnecker. 1987. Effects of
microwave treatment of live oak acorns on germination and on Curculio sp.
(Coleoptera: Curculionidae) larvae. J. Econ. Entomol. 80: 916-920.
D'Ambrosio, G., G. Ferrara & A. Tranfaglia. 1982. Teratogenesis due to
microwaves in Tenebrio molitor (Coleoptera: Tenebrionidae) and Sitophilus
granarius (Coleoptera: Curculionidae). Influence of impulse modulation.
Boll. Lab. Entomol. Agrar. Filippo Silvestri 39: 3-9.
Del Estal, P., E. Vinuela, E. Page & C. Camacho. 1986. Lethal effects of
microwaves on Ceratitis capitata Wied. (Diptera: Trypetidae) Influence of
developmental stage and age. J. Appl. Entomol. 102: 245-253.
Dost, W. A., W. W. Wilcox & A. N. Brier. 1987. California structural pest
control: a diverse industry. Voice of the PCOC, Fall 1987: 10, 24.
Ebeling, W. 1983. The Extermax system for control of the western drywood
termite, Incisitermes minor. Etex Ltd., Las Vegas. 11 p.
Ebeling, W. 1990. Heat and boric acid: an example of synergism. Pest
Control Technology April 1990: 44-46.
Ebeling, W. 1994. The Thermal Pest Eradication System for structural pest
control. The IPM Practitioner 16(2): 1-7.
Ebeling, W. & C. F. Forbes. 1988. Sand barriers for subterranean termite
control. The IPM Practitioner 10(5): 23-28.
Ebeling, W. & R. E. Wagner. 1964. Built-in pest control. Pest Control
32(2): 20-32.
Forbes, C. A. & W. Ebeling. 1986. Update: liquid nitrogen controls
drywood termites. The IPM Practitioner 8(8): 1-4.
Forbes, C. A. & W. Ebeling. 1987. Update: use of heat for elimination of
structural pests. The IPM Practitioner 9(8): 1-5.
Forest Products Laboratory. 1987. Wood handbook: Wood as an engineering
material. USDA Forest Service Agric. Handb. 72; rev. 1987. 466 p.
Hall, A. V. 1988. Pest control in herbaria. Taxon 37(4): 885-907.
Hall, D. W. 1981. Microwave: a method to control herbarium insects.
Taxon 30(4): 818-819.
Haverty, M. I., J. P. La Fage & W. L. Nutting. 1974. Seasonal activity
and environmental control of foraging of the subterranean termite,
Heterotermes aureus (Snyder), in a desert grassland. Life Sciences 15:
1091-1101.
Heinrich, B. 1981. Insect Thermoregulation. John Wiley and Sons, New
York, NY.
Kenaga, E. E. 1957. Some biological, chemical and physical properties of
sulfuryl fluoride as an insecticidal fumigant. J. Econ. Entomol. 50(1):1-6.
Kofoid, C. A. 1934. Climatic factors affecting the local occurrence of
termites and their geographical distribution, pp. 13-21. In C. A. Kofoid
[ed.], Termites and Termite Control. Univ. Calif. Press, Berkeley.
Kramer, R. 1992. The methyl bromide controversy heats up. Pest
Management 11(7): 26.
La Fage, J.P., M. I. Haverty & W. L. Nutting. 1976. Environmental factors
correlated with the foraging behavior of a desert subterranean termite,
Gnathamitermes perplexus (Banks) (Isoptera:Termitidae). Sociobiology 2(2):
155-169.
Levenson, H. & G. W. Frankie. 1983. A study of homeowner attitudes and
practices toward arthropod pests and pesticides in three U.S. metropolitan
area, pp 67-106. In G. W. Frankie and C. S. Koehler [eds], Urban
Entomology: Interdisciplinary Perspectives. Praeger, New York.
Lewis, V. R. & R. L. Lemaster. 1991. The potential of using acoustical
emission to detect termites within wood, pp. 34-37. In M.I. Haverty & W.W.
Wilcox [tech. coords] Proceedings of the Symposium on Current Research on
Wood-Destroying Organisms and Future Prospects for Protecting Wood in Use,
September 13, 1989, Bend, Oregon. USDA Forest Service General Technical
Report PSW-128.
Lewis, V. R., R. L. Lemaster, F. C. Beall & D. L. Wood. 1991. Using AE
monitoring for detecting economically important species of termites in
California. International Research Group on Wood Preservation, Document
IRG/WP/2375.
Locatelli, D. P. & S. Traversa. 1989. Microwaves in the control of rice
infestants. Ital. J. Food Sci. No. 2: 53-62.
Lund, A. 1962. Subterraneans and their environment. Pest Control 30:
3034, 36, 60-61.
Mampe, C.D. 1990. Termites, p. 201-262. In A. Mallis [ed.], Handbook of
Pest Control. Franzak & Foster Co, Cleveland, Ohio.
Meikle, R. W., D. Stewart & O. A. Globus. 1963. Drywood termite
metabolism of Vikane fumigant as shown by labeled pool technique. J.
Agraic. Food Chem. 11(3): 226-230.
Nelson, W. O. 1977. Microwave di electric properties of insects and grain
kernels. J. Microwave Power 12: 39.
Nelson, S. O. & J. A. Payne. 1982. Pecan weevil Curculio caryae control
by di electric heating. J. Microwave Power 17: 51-56.
Osbrink, W. L. A., R. H. Scheffrahn, N.-Y. Su & M. K. Rust. 1987.
Laboratory comparisons of sulfuryl fluoride toxicity and mean time of
mortality among ten termite species (Isoptera: Hodotermitidae,
Kalotermitidae, Rhinotermitidae). J. Econ. Entomol. 80(5): 1044-1047.
Osbrink, W. L. A., R. H. Scheffrahn, R.-C. Hsu & N.-Y. Su. 1988. Sulfuryl
fluoride residues of fumigated foods protected by polyethylene film. J.
Agric. Food Chem. 36(4): 853-855.
Philbrick, C. T. 1984. Comments on the use of microwave as a method of
herbarium insect control: possible drawbacks. Taxon 33(1): 73-74.
Reagan, B. M., J.-H. Chiao-Cheng & N. J. Streit. 1980. Effects of
microwave radiation on the webbing clothes moth, Tineola bisselliella and
textiles. J. Food Prot. 43(8): 658-663.
Rosenberg, U. & W. Bögl. 1987. Microwave pasteurization, sterilization,
blanching, and pest control in the food industry. Food Technology. June
1987: 92-99.
Rust, M.K., E. O. Paine & D. A. Reierson. 1995. Laboratory evaluation of
low temperature for controlling drywood termites. (In preparation).
Rust, M. K., J. K. Grace, D. L. Wood & D. A. Reierson. 1988. The search
for new termite control strategies. Calif. Agric. 42(5): 15-18.
SAS Institute. 1994. SAS/STAT guide for personal computers, version 6.10
ed. SAS Institute, Cary, NC.
Scheffrahn, R. H., W. L. A. Osbrink, R.-C. Hsu & N.-Y. Su. 1987.
Desorption of residual sulfuryl fluoride from structural and household
commodities by headspace analysis using gas chromatography. Bull. Environ.
Contam. Toxicol. 39: 769-775.
Scheffrahn, R. H., R.-C. Hsu, W. L. A. Osbrink & N.-Y. Su. 1989a.
Fluoride and sulfate residues in foods fumigated with sulfuryl fluoride.
J. Agric. Food Chem. 37: 203-206.
Scheffrahn, R. H., R.-C. Hsu & N.-Y. Su. 1989b. Fluoride residues in
frozen foods fumigated with sulfuryl fluoride. Bull. Environ. Contam.
Toxicol. 43: 899-903.
Scheffrahn, R. H., R.-C. Hsu & N.-Y. Su. 1990. Evaluation of polymer film
enclosures as protective barriers for commodities from exposure to
structural fumigants. J. Agric. Food Chem. 38(3): 904-908.
Scheffrahn, R. H. and N.-Y. Su. 1992. Comparative toxicity of methyl
bromide against ten nearctic termite species (Isoptera: Termopsidae,
Kalotermitidae, Rhinotermitidae). J. Econ. Entomol. 85(3): 845-847.
Scheffrahn, R. H., L. Bodalbhai, & N.-Y. Su. 1992. Residues of methyl
bromide and sulfuryl fluoride in manufacturer-packaged household foods
following fumigation. Bull. Environ. Contam. Toxicol.48: 821-827.
Scheffrahn, R. H., W. P. Robbins, P. Busey, N.-Y. Su & R. K. Mueller.
1993. Evaluation of a novel, hand-held, acoustic emissions detector to
monitor termites (Isoptera: Kalotermitidae, Rhinotermitidae) in wood. J.
Econ. Entomol. 86(6): 1720-1729.
Scheffrahn, R. H., G. S. Wheeler, and N.-Y. Su. 1995. Synergism of methyl
bromide and sulfuryl fluoride toxicity against termites (Isoptera:
Kalotermitidae, Rhinotermitidae) by admixture with carbon dioxide. J.
Econ. Entomol. 88(3): 649-653.
Scheffrahn, R. H., C. L. Bloomcamp & N.-Y. Su. 199?. Indoor airborne
residues of methyl bromide and sulfuryl fluoride following aeration of
fumigated houses. (In review/press??).
Smythe, R. V. & L. H. Williams. 1972. Feeding and survival of two
subterranean termite species at constant temperature. Ann. Entomol. Soc.
Am. 65(1): 226-229.
Steel, R. G. F. & J. H. Torrie. 1960. Principles and Procedures of
Statistics. McGraw-Hill Book Company, Inc. New York.
Su, N.-Y. & R. H. Scheffrahn. 1986. Field comparison of sulfuryl fluoride
susceptibility among three termite species (Isoptera: Kalotermitidae,
Rhinotermitidae) during structural fumigation. J. Econ. Entomol. 79:
903-908.
Su, N.-Y. & R. H. Scheffrahn. 1990. Economically important termites in
the United States and their control. Sociobiology 17: 77-94.
Tallon, J. C. 1992. Non-toxic method of exterminating insects. U. S.
Patent 5,165,199.
Thoms, E. M. & R. H. Scheffrahn. 1994. Control of pests by fumigation with
Vikane gas fumigant (sulfuryl fluoride). Down to Earth 49(2): 23-30.
Tilton, E. W. & H. H. Vardell. 1982a. An evaluation of a pilot-plant
microwave vacuum drying unit for stored-product insect control. J. Ga.
Entomol. Soc. 17(1): 133-138.
Tilton, E. W. & H. H. Vardell. 1982b. Combination of microwaves and
partial vacuum for control of four stored-products insects in stored grain.
J. Ga. Entomol. Soc. 17(1): 106-112.
Watters, F. L. 1976. Microwave radiation for control of Tribolium
confusum in wheat and flour. J. Stored Prod. Res. 12: 19-25.
Wilcox, W. W. 1979. Decay of wood in structures. Calif. Agric. 33: 32-33.
Figure Captions
 
Figure 1. Tongue depressor holding chambers used for maintaining groups of
Incisitermes minor prior to testing or after testing to measure 4-wk
mortality.
Figure 2. Villa Termiti used to test fumigation techniques and nonchemical
alternatives to fumigation for control of drywood termites. The Villa
Termiti is symmetrical with all four sides structurally identical.
Figure 3. Sagittal view of Villa Termiti showing test board placement
locations for attic, drywall, and subarea.
Figure 4. Sagittal view of foundation showing test board placement
locations for tests on or near the sill plate.
Figure 5. Douglas-fir "2 by 4" cut so that artificial infestations of 25
I. minor pseudergates and nymphs could be placed in each of three galleries.
Figure 6. Thermocouple placement for attic, drywall, and subarea treated
by excessive heat.
Figure 7. Schematic drawing of drywall area in Villa Termiti. Shaded
circles indicate points of introduction for liquid nitrogen. Location of
thermocouples are indicted by arrows.